System Options:
Heating Stage 1000 °C / Heating Stage 1500 °C
9-43
150 - Heating Stage Settings
This is the supervisor alignment procedure.
•
Step 1
: the
Temperature Ramping Limit
could be set up to 300
K/min (default value is 50 K/min).
Caution!
Ramp speed higher than 100 K/min strongly decreases the heater
lifetime and leads to regulation instability.
•
Step 2
(applicable for HS 1500 only): highly advanced setting
intended for practised HS users.
Содержание Quanta FEG 250