System Options:
Heating Stage 1000 °C / Heating Stage 1500 °C
9-42
The Advanced tab
The Advanced tab is displayed only when the HS is connected to the
feed-through plate connector.
•
The
Auto Power
check box is cleared, the
Power
slider becomes
active and user can apply desired power directly to the heater. The
functionality is useful when working at temps above 1300 °C, when
regulation can be affected by a sample outgassing or a leakage
current.
Caution!
The slider button shows % of maximum allowed power. Apply just
as many Watts as needed in order not to exceed 50 °C/min ramp
speed, otherwise your heater lifetime shortens.
For advanced
users ONLY.
Additional HS 1500 features
•
The
Bias Presets
check box allows the user to apply sample and
shield bias values according to settings in the alignment procedure
150 (see below).
•
The
Heat Shield
and
Sample Bias
sliders are used for the manual
sample and shield bias values setting.
•
The power and biases can also be controlled via
Manual User
Interface
(MUI) modified knobs, which is a part of the 1500 °C
heating stage option. The MUI performance could be set in the
Preferences…
/
General
tab, where the following line appears:
MUI knobs assignment
(Default / Heating stage)
The Default setting always keeps the original functionality. The
Heating Stage setting assigns MUI knobs another functionality
when the
Temperature Control
module /
Heating
button is
clicked. When the HS is turned off or removed, the habitual
functionality is restored (see above).
FIGURE 9-36
MUI BUTTONS NEW FUNCTIONALITY
•
the
Stigmator X
knob change to the
Power
knob
•
the
Stigmator Y
knob change to the
Enhanced Contrast
knob
•
the
Shift X
knob changes to the
Sample Bias
knob
•
the
Shift Y
knob changes to the
Heat Shield Bias
knob
Содержание Quanta FEG 250
Страница 8: ...C vi...
Страница 28: ...System Operation Quanta FEG System States 3 10...
Страница 108: ...Alignments 1 Gun Alignment 6 6...
Страница 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Страница 152: ...System Options Automatic Aperture System 9 4...
Страница 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Страница 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...