Alignments:
5 - Emitter Startup
6-10
1.
Check the
Gun vacuum page:
if it is
IGP ON
click the Next button.
If it is
OFF
click the
IGP On
button to automatically start the ion
pump. This procedure takes a few minutes. If the ion pump start is
not successful, call service.
2.
Click the
Emitter On / Off
button to start / stop the emission of
electrons.
Note:
If the Emitter On button is disabled and the IGP´s are running,
restart the xT Microscope Server and try again.