Maintenance and Troubleshooting:
Aperture Strip Module
8-8
REPLACING THE APERTURE MODULE
The new Aperture Module comes in a container, has been cleaned
and is ready to be fitted to the rod.
1.
Unscrew the Titanium screw holding the old module onto the rod.
Keep the screw in the hole of the rod and let the module fall away.
2.
Open the new module pack and let the new module sit with the
connection end uppermost to the edge of the container base.
3.
Pick up the new module with the Titanium screw end and fasten,
making sure of a good fit.
REPLACING THE APERTURE ROD
1.
Check that there are no fibres on the rod ‘o’ ring. Do not grease the
‘o’ ring.
2.
Replace the Aperture rod back into the Aperture Adjuster
assembly on the column and turn the end screw mechanism until
the holder is hand tight.
3.
Pump the microscope specimen chamber.
4.
Reconnect the heater cable to the outer end of the rod if
necessary.
5.
Select the 30 µm aperture so that alignment can be performed.