System Options:
Plasma Cleaner
9-56
104 - PLASMA CLEANING ALIGNMENT
When repeated
Sample Cleaning…
procedure is not efficient (see
above), click the
Start Chamber Cleaning
button, (the procedure
Duration
is much longer).
Warning!
Do not use Chamber cleaning procedure when EDS / WDS / EBSD
system is mounted! If you need to clean system with EDS detector
mounted, use 1 mm collimator and do not exceed 30 minutes once a
week at most.