System Operation:
Quanta FEG System States
3-7
Quanta FEG System States
There are several system states:
0. Complete Shutdown
– service and emergency reasons
1. Shutdown
– when not using the system for more than 10 days
2. Standby
– when not using the system for more than 1 day
3. Overnight
– when not using the system overnight
4. Full Operation
– when working
Note:
a)
Once you have your
FEI Microscope user
(or
Supervisor
)
account
set up via
FEI User management
software by
FEI Account
Administrator
(see Chapter 4), you can use your name and
password to access both Windows xP system and the
xT microscope Control
software. (Take note of the case sensitive
passwords necessary at Windows xP and xT microscope Control
server Log On points. A password is advisable for logging on to
protect individual settings and results.)
TABLE 3-2
STARTUP PROCEDURE GENERALLY
System State Action
0. Complete
Shutdown
1)
Connect the power cord to the microscope
console, a compressed air and a nitrogen inlet
(optional) for venting and cooling. Interlocks
prevent the vacuum system from operating if any
of these are not present (with the exception of
nitrogen inlet).
1. Shutdown
2)
Push the
power
button on the microscope front
control panel.
2. Standby
3)
Switch on the PC. The operating system
(a)
Windows xP loads and displays the appropriate
icons on the monitor desktop.
4)
Double-click the
xT microscope Server
icon to
start the software (all seeming LEDs should be
green).
5)
Click the
Start
icon to start the server. Wait until
all dialogues are fully functional.
3. Overnight
6.
Click the
Start UI
button to start the
xT
microscope Control
software. The main window
appears behind the
XTUI Log On
dialogue.
7.
Enter your
Username
and
Password
(a)
. When
the xT microscope Control software is available,
the
Beam Control
page displays on screen.
8.
Switch on the Emitter (see Chapter 6).
9.
Click the
Vacuum
module /
Pump
Button. Wait
for the
Pumped
status.
10.
Click the
Column
module /
Beam On
button.
(must be yellow).
4. Full
Operation
11.
Select a quad, a detector and beam
parameters and resume the quad.
Содержание Quanta FEG 250
Страница 8: ...C vi...
Страница 28: ...System Operation Quanta FEG System States 3 10...
Страница 108: ...Alignments 1 Gun Alignment 6 6...
Страница 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Страница 152: ...System Options Automatic Aperture System 9 4...
Страница 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Страница 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...