Maintenance and Troubleshooting:
Troubleshooting
8-14
Poor ETD image
•
Select the ETD. Open the
Detectors
menu /
Detector
Settings
. Select the
Mode:
Custom
. Change the
Grid
Voltage
from maximum to minimum (see Chapter 5).
•
There should be a visible change of the contrast (an
image is getting darker with lower voltage). If not, run
the
Diagnostics Auto Report
and
Simple TAD
utilities
(see below).
Alternating lighter and darker image
bands visible in the ESEM / LoVac mode
•
Run the
Purging
procedure (see Chapters 3 and 4).
•
Check the water bottle to be sealed properly (see above).
Discharging and white stripes
in ESEM / LoVac mode
•
Decrease the
Detector
module /
Contrast
value (see
Chapter 4).
Water interlock failure
•
When cooling water does not flow, the
Application
Status…
window with the respective message pops up
(see Chapter 4). Check the cooling water flow.
Note:
The system is designed to work without the cooling water.
However, if the air speed near the column is higher than 5
m/min, it may cause an image drift due to the temperature
fluctuation.
Compressed air interlock failure
•
If the compressed air is not supplied or its pressure is
too low, the
Application Status…
window with the
respective message pops up (see Chapter 4). Check
the compressed air pressure at the microscope input
(either on compressor or at the local air distribution).
The input pressure must be within the range:
5.5 – 8.0 bar (550 – 800 kPa).
Note:
The compressed air is essential for the system, which
cannot work without it.
Emission lost after power failure
•
Run the
5 - Emitter Startup
procedure to start the
emission (see Chapter 6).
Emission is missing
•
Run the
5 - Emitter Startup
procedure to start the
emission (see Chapter 6). If this is not possible, restart
the whole system and try again.
•
If this doesn’t help, run the
Diagnostics Auto Report
and
Simple TAD
utilities (see below).
Specimen holder wizard doesn’t
proceed
•
Read the
Specimen Holder Wizard
procedure for the
possible reasons (stub holder height, manual tilt stage
tilt angle readout) (see Chapter 7).
Pressure doesn’t reach the target value
when using the water in the ESEM /
LoVac mode
•
Check the water bottle to be sealed properly, to be filled
with the water and the heating plate is warm (see
above).
Frozen system
•
Restart the PC.
Touch Alarm is active
•
Remove the specimen holder away from the chamber.
•
Check the presence of a short connection between the
specimen holder and the stage.
Other problems
•
Run the
Diagnostics Auto Report
and
Simple TAD
utilities (see below).
TABLE 8-2
POSSIBLE PROBLEMS AND CORRECTIVE ACTIONS
Содержание Quanta FEG 250
Страница 8: ...C vi...
Страница 28: ...System Operation Quanta FEG System States 3 10...
Страница 108: ...Alignments 1 Gun Alignment 6 6...
Страница 114: ...Alignments 154 Water Bottle Venting 6 12 154 Water Bottle Venting...
Страница 152: ...System Options Automatic Aperture System 9 4...
Страница 154: ...System Options Nav Cam 9 6 401 NAV CAM AUTOBRIGHTNESS...
Страница 177: ...System Options Cooling Stage Waterless Cooling Stage 9 29...