7.9.4
Strip and separation chamber
This section describes how to manually clean the strip and separation chamber and replace the glass
bead.
a0012478
Fig. 55:
Position of the strip and separation chamber
You require the following parts:
• 4 mm Allen key
• Injector (part of the scope of supply)
• Glass bead
• Soft brush
• Paper towels to absorb liquid
• Vessel with volume of approx. 150 ml (5 fl.oz) to collect liquid
• Needle-nosed pliers