TOCII CA72TOC
Maintenance
Hauser
73
a0014230
Fig. 36:
Pump P2
Pump P2
: Pumps sample from the separation chamber into the furnace.
Hose: (BK-BK) 0.76 mm (0.03")
1
2
3
4
5
6
1
Pump retainer
4
Adjusting screw
2
Pump head
5
Hose cassette
3
Sample hose (BK-BK) 0.76 mm (0.03") ID
6
Motor
Step
Display
Description
1
2
#
Warning!
There is a risk of infection when working
with wastewater! For this reason, wear
protective goggles, protective gloves and
protective clothing.
Press the [E] key.
PUMPS
CALIBRATION
S E R V I C E
CLEANING
FILTERS
P 1/4 REPLACE HOSE
PUMP 2 REPLACE HOSE
PUMP 3 REPLACE HOSE
CALIBRATE PUMP 2
REPLACE HOSE P2
CAUTION! LIQUID CAN LEAK OUT
WEAR PROTECTIVE CLOTHES
[E] TO CONFIRM