TOCII CA72TOC
Operation
Hauser
31
a0012449-en
Fig. 12:
Spot view (example)
1
Time mark on the load curve
2
Measured value of the selected time
You can switch the spot view (cursor) off with the Enter key.
5.4
Standby mode
The standby mode can be used to operate the analyzer at measuring points where the flow of sample
is interrupted at times.
The option is available in the one-channel version with a PA-2 or PA-3 sample conditioning system.
a0013853
Fig. 13:
Standby switch
1
Adapter fitting for the pressure monitor
2
Pressure monitor ¼"
3
Signal connection
4
Protection cap
5
Bypass screen
Mode of operation
If the flow of sample is interrupted, the pressure monitor reports this status to the computer (DI = 1)
via the switch input DI5.
This causes the device to respond as follows:
• All the pumps are stopped.
• Pump P2 is drained.
• The strip chamber is rinsed.
• The current output is set to 0.0 mA.
• The analyzer operates without displaying a measured value.
The device resumes measuring operation automatically as soon as sample flows again.
5
00
6
00
7
00
8
00
9
00
10
00
11
00
1000
500
1
2
238 mg/l
8:30
RECORD
DATE 08/29/08
VALUE=
TIME=
V
1 hour earlier
U
2 minutes earlier
W
1 hour later
T
2 minutes later
1
2
3
4
5