Maintenance
Hauser
77
3
1.
Release the hose connection from
pump P1 to the strip chamber.
2.
Have a vessel at the ready to collect
the liquid.
3.
Drain the strip chamber.
4.
Press the [E] key.
4
Caution! Acid!
1.
Remove the acid suction hose from
the acid cistern and place the end in
a collecting vessel.
2.
Release the hose cassette.
3.
Make sure that the acid runs back
into the collecting vessel.
!
Note!
No medium containing TOC should enter
the acid feeder!
4.
Press the [E] key.
5
In areas with a low amount of dust:
Grease the new hoses with a light coating
of silicone grease.
1.
Release the pump hose from the
connections.
2.
Change the hose.
3.
Press the [E] key.
6
1.
Connect the hose of pump P1 to the
strip chamber.
2.
Press the [E] key.
Step
Display
Description
REPLACE HOSE P3
[E] TO CONFIRM
EMPTY STRIP CHAMBER BY
DISCONNECTING HOSE CONNECTION OF
SAMPLE PUMP P1 AT THE STRIP CHAMBER
PLACE VESSEL BELOW TO COLLECT
CONTENT OF STRIP CHAMBER
REPLACE HOSE P3
[E] TO CONFIRM
DANGER! ACID
REMOVE ACID SUCTION HOSE FROM
ACID FEEDER AND PLACE VESSEL
BELOW TO COLLECT ACID
RELEASE HOSE CASSETTE P3 AND EMPTY
ACID LINE
REPLACE HOSE P3
[E] TO CONFIRM
REPLACING PUMP HOSE
REPLACE HOSE P3
[E] TO CONFIRM
CONNECT SAMPLE HOSE PUMP P1
TO STRIP CHAMBER