TOCII CA72TOC
Commissioning
Hauser
33
6.2.2
CLR start
You must perform a CLR start the first time you put your analyzer into operation. This loads the
analyzer software and deletes all the previous data saved. The parameters remain.
Perform a CLR start as follows:
1.
Press and hold down the CLR key for approx. 5 seconds while switching on the main switch
simultaneously.
The application is loaded.
2.
The following information appears on the screen once the program has been loaded:
3.
Press the [E] key.
4.
The following information appears on the screen:
5.
Press the [E] key.
The analyzer heats up the reactor. Measuring operation starts automatically as soon as the set
temperature is reached.
You have to adapt the analyzer to the conditions of the measuring point following the CLR start.
Refer to the "Optimizing the analyzer" section for this purpose. Then adjust the setting data of your
analyzer in the "Programming mode".
CA72TOC-xxxxx
VERS. xxx
MAINTENANCE COMPLETED?
[E] TO CONFIRM
Содержание TOCII CA72TOC
Страница 2: ...Endress Hauser...
Страница 136: ...TOCII CA72TOC 136 Endress Hauser...
Страница 138: ...www endress com worldwide BA448C 07 en 06 10 71118197 Printed in Germany FM SGML 6 0 DT 71118197...
Страница 139: ......
Страница 140: ......