Commissioning
TOCII CA72TOC
44
Hauser
6.3.12
Menu TEST - TEST OF OUTPUTS
Step
Display
Description
1
2
1.
Select the desired outputs.
2.
Press the [E] key.
3
Current output
Sets the analog current outputs to any
value between 0 and 20 mA.
Pumps
Parameter for testing the function of the
pumps
Switch outputs
Displays the switching status of the
switch outputs (see the following table).
SETTING
LISTS
P R O G R A M M I N G
TEST
TEST OF INPUTS
TEST OF OUTPUTS
TEST FREQUENCY I/O
TEST COM2
MEASUREMENT OFF
TEST
O U T P U T S
DC-SIGNAL
DIGITAL OUTPUTS
PUMPS
0.0 mA
0.0 mA
TEST
DC-SIGNAL
CHANNEL NO. 1
CHANNEL NO. 2
xxx.x %
xxx.x %
xxx.x %
xxx.x %
xxx.x ml
xxx.x ul
xxx.x ul
xxx.x ml
TEST
PUMPS
PUMP NO. 1
PUMP NO. 2
PUMP NO. 3
PUMP NO. 4
PUMP NO. 5
PUMP NO. 1/2
ON
TEST
DIGITAL OUTPUTS
ON / OFF : USE ANY KEY
SWITCH CHANNEL : USE ARROW KEY
DO-NO. 1
DO-NO. 2
DO-NO. 3
DO-NO. 4
DO-NO. 5
DO-NO. 6
DO-NO. 7
DO-NO. 8
OFF
ON
OFF
OFF
OFF
ON
OFF
OFF
DO-NO. 9
DO-NO. 10
DO-NO. 11
DO-NO. 12
OFF
ON
OFF
OFF