Commissioning
TOCII CA72TOC
46
Hauser
6.3.15
Menu TEST - MEASUREMENT OFF
a0012618-en
Fig. 27:
"Test" menu
Deactivates measuring operation. The MEASUREMENT OFF operating status is displayed. Select
this status to perform tests without triggering an alarm.
6.3.16
Two-channel: external changeover
The analyzer is equipped with one or two separate sample supply systems.
The current sample selected is controlled externally by signal input 8 (binary in 8).
Signal input 8 = 0
channel 1
Signal input 8 = 1
channel 2
If the analyzer is operated with
one
sample conditioning system, the operator must ensure that the
right sample is at the bypass when a channel switch is requested.
If the analyzer is operated with
two
sample conditioning systems, the solenoid valve MV6 is used
internally to switch the channels.
If the signal status at signal input 8 changes, the measuring cycle is terminated immediately and the
system immediately commences channel switchover.
If the "Operation" key is pressed during channel switchover, the process of switching channels is
aborted and the measuring cycle commences in the active channel. Conditioning of the analyzer to
the sample in the active channel is suppressed.
!
Note!
The measuring channel cannot be switched manually.
Settings for the graphic interface
You must enter the following parameters in the Range Data menu:
PROGRAMMING/SETTING/RANGE DATA:
SETTING
LISTS
P R O G R A M M I N G
TEST
TEST OF INPUTS
TEST OF OUTPUTS
TEST FREQUENCY I/O
TEST COM2
MEASUREMENT OFF
Parameter
Unit
Description
SCALE CH1
mg TOC/l
Use this function to enter the maximum concentration for channel 1
here. This value determines the scale end value for the graphic
interface.
SCALE CH2
mg TOC/l
Use this function to enter the maximum concentration for channel 2
here. This value determines the scale end value for the graphic
interface.
Содержание TOCII CA72TOC
Страница 2: ...Endress Hauser...
Страница 136: ...TOCII CA72TOC 136 Endress Hauser...
Страница 138: ...www endress com worldwide BA448C 07 en 06 10 71118197 Printed in Germany FM SGML 6 0 DT 71118197...
Страница 139: ......
Страница 140: ......