Maintenance
TOCII CA72TOC
104
Hauser
Step
Display
Description
1
2
Clean the needle as follows:
1.
Remove the hose P2 from the
capillary of the dosing head.
2.
Release the red screw plug (item 1).
3.
Remove the dosing head (item 5).
4.
Using a damp cloth, remove the salt
residue on the capillary or replace the
capillary.
5.
Check the O-ring (item 4), and also
check the O-ring (item 3) if replacing
the capillary.
6.
Insert the dosing head and tighten
the red screw plug.
7.
Fit the hose P2 onto the capillary.
8.
Press the [E] key.
Operation starts.
PUMPS
CALIBRATION
S E R V I C E
CLEANING
FILTERS
SCREEN FLUSH
POWER-FLUSH
BYPASS-SCREEN
SSEPARATION
OPEN GAS CICUIT
COMBUSTION PIPE
LEAKAGE TEST
OPENING THE GAS CIRCUIT
[E] TO CONFIRM
MAINTENANCE COMPLETED ?