Maintenance
TOCII CA72TOC
114
Hauser
Step
Display
Description
1
2
#
Warning!
Wear protective gloves!
1.
Release the couplings (items 1 and
4).
2.
Remove the acid filter (item 3) from
the retaining clips (item 2).
3.
Remove the filling in the acid filter.
4.
Clean the glass if necessary.
5.
Roll the fabric into a roll and push it
into the filter glass (item 8).
!
Note!
Do not press it in too hard. Shorten the
fabric if necessary.
6.
Fill the glass up to the halfway mark
with copper and then with zinc (see
order of items 6 and 7). Allow
enough room for the second piece of
fabric.
7.
Roll the fabric into a roll and use it to
close the filling in the acid filter (item
5).
8.
Clean the O-rings and seal the acid
filter. Make sure that the fabric does
not extend as far as the plug to
ensure the filter housing is sealed
properly (see zoom detail of
Abb. 63).
9.
Fit the acid filter into the retaining
clips and connect the filter.
10. Press the [E] key.
Operation starts (initially without a
measured value).
PUMPS
CALIBRATION
S E R V I C E
CLEANING
FILTERS
REPLACE ACID FILTER
REPLACE GAS FILTERS
REPLACE O2 FILTERS
REPLACE HEATED FILTER
REPLACE ACID FILTER
[E] TO CONFIRM
MAINTENANCE COMPLETED ?