Commissioning
TOCII CA72TOC
52
Hauser
6.5
Communication
The analyzer is equipped with an RS-232 serial interface.
Data transmission is unidirectional and is performed with the following parameters:
The string is 104 bytes long and is sent every 2 seconds.
Baud rate
9600 baud
Bits
8 bit
Parity
N
Stop bit
1 bit
Handshake
No
Byte
Description
0
Start byte
1
0 = Operation disabled
1 = Operation enabled
2
0 = Emergency shutdown
1 = Channel 1 measuring operation enabled
2 = Adjustment or calibration
3 = Service
4 = Programming
5 = Channel 2 measuring operation enabled
3
Leakage (0 = off, 1 = on)
4
Temperature too high (0 = off, 1 = on)
5
Low carrier gas supply (0 = off, 1 = on)
6
IR detector fault (0 = off, 1 = on)
7
Temperature too low (< 85 % T
set
) (0 = off, 1 = on)
8
Outside the measuring range (0 = off, 1 = on)
9
Temperature deviation of Peltier cooler (T
set
± 3 °C) (0 = off, 1 = on)
10
pH alarm (0 = off, 1 = on)
11
Temperature deviation (< T
set
-30 °C) (0 = off, 1 = on)
12
Standby (0 = off, 1 = on)
13
Limit value exceeded (0 = off, 1 = on)
14
Limit value undershot (0 = off, 1 = on)
15
Slope alarm (0 = off, 1 = on)
16
Unstable dosing, sample failure (furnace) (0 = off, 1 = on)
17
Water supply failure (0 = off, 1 = on)