TOCII CA72TOC
Maintenance
Hauser
111
7.9.7
Leakage test
This section describes how the gas circuit is checked for leaks.
Check the leak-tightness of the gas circuit after each modification to the furnace.
You require the following parts:
• Drain plug from the accessories supplied
a0012531
Fig. 61:
Left-hand side of the housing
1
Gas outlet
2
Switch for membrane compressor
1
2
Step
Display
Description
1
2
1.
Switch off the membrane compressor
(item 2).
2.
Seal the gas outlet with a plug
(left-hand side panel, item.1).
3.
Press the [E] key.
PUMPS
CALIBRATION
S E R V I C E
CLEANING
FILTERS
SCREEN FLUSH
POWER-FLUSH
BYPASS-SCREEN
SSEPARATION
OPEN GAS CICUIT
COMBUSTION PIPE
LEAKAGE TEST
LEAKAGE TEST
[E] TO CONFIRM
SHUT GAS OUTLET
SWITCH MEMBRANE COMPRESSOR OFF