TOCII CA72TOC
Commissioning
Hauser
49
All the alarm limits affect the same signal output II (binary out II). A limit value alarm is also retained
following channel switchover until the limit value for the channel in question is undershot.
Interruption of time control
The duration of the measurement can be cancelled in the following way:
• By manual input
During measurement operation with the
keys 1
and
2
you can switch over to the other channel.
• Signal input binary in 8
binary in 8 = 0
no effect
binary in 8 = 1
(for about 10 seconds) switch over to the other channel
When you force a switch over to the other channel via keyboard or binary in 8 the measuring cycle
will be stopped immediately and the switch over to the other channel starts.
Parameter
Unit
Description
HI ALARM LIMIT CH1
mg/l
Use this function to specify the alarm value for when the value for
channel 1 is overshot.
LO ALARM LIMIT CH1
mg/l
Use this function to specify the alarm value for when the value for
channel 1 is undershot.
HI ALARM LIMIT CH2
mg/l
Use this function to specify the alarm value for when the value for
channel 2 is overshot.
LO ALARM LIMIT CH2
mg/l
Use this function to specify the alarm value for when the value for
channel 2 is undershot.