TOCII CA72TOC
Maintenance
Hauser
93
5
1.
Rinse the sensor with deionized
water.
2.
Hold the sensor in the buffer solution
pH = 7.00.
3.
Press the [E] key.
6
1.
Wait until the measured value
stabilizes (a bar appears to the right of
the measured value).
2.
Press the [E] key.
Calculation of the calibration values.
Typical slope value: between
55 mV/decade and 58 mV/decade
7
1.
Put the pH sensor into the strip
chamber.
2.
Screw the cover of the strip chamber
closed hand-tight.
3.
Press the [E] key.
Operation starts.
The following appears on the display if the adjustment failed:
Step
Display
Description
ADJUSTMENT pH SENSOR
[E] TO CONFIRM
RINSE SENSOR WITH DI WATER
PLACE SENSOR INTO SOLUTION pH= 7.00
X.XX
ADJUSTMENT pH SENSOR
[E] TO CONFIRM
MEASURED VALUE
ADJUSTMENT pH SENSOR
[E] TO CONFIRM
RE-INSERT pH-SENSOR
Содержание TOCII CA72TOC
Страница 2: ...Endress Hauser...
Страница 136: ...TOCII CA72TOC 136 Endress Hauser...
Страница 138: ...www endress com worldwide BA448C 07 en 06 10 71118197 Printed in Germany FM SGML 6 0 DT 71118197...
Страница 139: ......
Страница 140: ......