TOCII CA72TOC
Wiring
Hauser
23
4.1.3
Signal connection
a0011911
Fig. 6:
Signal connection
!
Note!
TOCII CA72TOC is suitable for connection to industrial mains support according to EN 61326-1,
class A.
binary
out
binary
out
1
2
3
4
5
6
7
8
binary
in
binary
in
0/4-20
mA
0/4-20
mA
RS 232
RS 232
I
III
IV
II
+
-
+
-
40
41
1
2
1
2
1
2
1
2
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
+24VDC
GND
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
I
Alert
1
Calibration, external activation
II
Alarm for limits
2
Adjustment, external activation
III
Standby
3
Screen flush, external activation
IV
Operating control
4
Power flush, external activation
40
Signal output channel 1
5
Not assigned
41
Signal output channel 2 (optional)
6
Not assigned
7
Standby, external activation
8
Channel switchover, external activation (optional)
Messages I to IV
Potential-free relay contact (max. 0.2 A and 50 V), normally closed
(NC)
Relay contact I closed = no fault messages
Relay contact II closed = no collective alarm
Relay contact III closed = no standby
Relay contact IV closed = operating check
At the end of a measuring cycle, relay IV opens for 2 seconds to
signal the end of the measuring cycle.
Signal outputs 40 to 41
0 to 20 mA or 4 to 20 mA can be switched, galvanically isolated,
max. load 500
Ω
Signal inputs 1 to 8
24 V DC active, max. load 500
Ω