Technical Data
TOCII CA72TOC
132
Hauser
10
Technical Data
10.1
Input
10.2
Output
10.3
Power supply
Measured variable
TOC (TC possible without stripping unit)
Measuring range
Version
Measuring range
A
0.25 to 600 mg TOC/l
B
1 to 2400 mg TOC/l
C
2.5 to 6000 mg TOC/l
D
5 to 12000 mg TOC/l
With optional predilution, the measuring range can be expanded by a factor of 20.
Signal inputs
8 signal inputs 24 V DC, active, load max. 500
Ω
Input #1
Service, trigger calibration
Input #2
Service, trigger adjustment
Input #3
Service, trigger screen flush
Input #4
Service, trigger power flush
Input #5
Not assigned
Input #6
Not assigned
Input #7
Trigger standby
Input #8
Trigger channel switchover (optional)
Output signal
Measuring channel 1: 0/4 to 20 mA, galvanically isolated
Measuring channel 2 (for two-channel version): 0/4 to 20 mA, galvanically isolated
Load: max. 500
Ω
Data interface
RS 232 C, proprietary, for data output and remote operation (optional)
Alarms
4 outputs: Limit value alarm, fault message, standby message and operating check,
potential-free, normally closed (max. 0.25 A / 50 V)
Supply voltage
115/230 V AC, 50/60 Hz, 650 VA
Fuses
Assembly
Fuses
Power distribution
2.5 A, slow-blow, design: fine-wire fuse 6.3 x 32
Relay
4 A per relay, slow-blow, design: TR5
Power unit
2 A, slow-blow, design: fine-wire fuse 5 x 20
!
Note!
TOCII CA72TOC is suitable for connection to industrial power supply systems in accordance with EN 61326-1, Class A.