Commissioning
TOCII CA72TOC
40
Hauser
6.3.9
Menu LISTS - ALARM RECORD
a0012587-en
Fig. 21:
"Lists" menu
All the alarm messages, including the date and time of the event, are logged in the alarm record.
a0012592-en
Fig. 22:
Alarm record
REPLACE GAS FILTERS
Date and time when selected in the service menu
REPLACE O2 FILTER
Date and time when selected in the service menu
HEATED FILTER
Date and time when selected in the service menu (optionally heated salt trap)
STANDBY
Date and time of a standby event
BASIC LINE DRIFT
Date and time of the basic line drift when calibrating and adjusting
Value 1: Basic line offset [ppm]
Value 2: Basic line drift slope [ppm/min]
EMPTY VOLUME DOSING
Date and time when selected in the service menu
Value 1: Duration of the filling process [s]
Value 2: Volume [
μ
l]
CALIBRATION
Date and time of the calibration of the analyzer and the TOC value found,
and the recovery with regard to the specified concentration of standard C2
Value 1: TOC [mg/l]
Value 2: Recovery [%]
SET DEFAULTS
Date and time when selected in the Setting menu
Action
Description
SETTING
LISTS
P R O G R A M M I N G
TEST
MAX MIN AVERAGE
MAINTENANCE RECORD
SHOW COMPLETE RECORDS
ALARM RECORD
ALARM RECORD
XX.XX
XX:XX
XXXXXXXXXXXXX
XXX
Date
Time
Event
on/off
XX.XX
XX:XX
XXXXXXXXXXXXX
XXX