Replacing the hose on pump P2
This section describes how the hose of pump P2 is replaced. The following service menus are
automatically started when this service menu finishes:
1.
Adjusting pump P2
2.
Empty volume dosing
#
Warning!
Danger of crushing! Never reach into the pump head while the pump is in operation!
a0012484
Fig. 35:
Position of pump P2
You require the following parts:
• 10 ml graduated cylinder
• 2.5 mm Allen key
• Dosing nozzle (injector)
• Paper to absorb liquids
• Collecting vessel with volume of approx. 150 ml (5 fl.oz) to collect liquid
• Silicone grease