If it is not, correct the problem and reboot or power-cycle the transmitter.
•
If it is, replace the transmitter because the transmitter has an internal power issue.
17.13.26 [104] Calibration in Progress
Alert
Function Check in Progress
Cause
A calibration is running.
Recommended actions
Allow the test to complete.
17.13.27 [105] Two-Phase Flow
Alert
Process Aberration
Cause
The line density is outside the user-defined two-phase flow limits.
Recommended actions
1. Check for two-phase flow.
Configure Lower Range Value (LRV) and Upper Range Value (URV) for the mA Output
2. Verify that the transmitter is configured correctly for the connected sensor.
17.13.28 [110] Frequency Output
X
Saturated
Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the scaling of the Frequency Output.
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
Troubleshooting
Configuration and Use Manual
March 2019
MMI-20025166
272
Micro Motion Model 5700 Transmitters with Configurable Outputs