17.13.36 [119] Discrete Output 2 Fixed
Alert
Output Fixed
Cause
The output is set to a constant state. A loop test may be in progress.
Recommended actions
Check to see if a loop test is in process (the output is fixed).
17.13.37 [120] Curve Fit Failure (Concentration)
Alert
Configuration Error
Cause
The transmitter was unable to calculate a valid concentration matrix from the current data.
Recommended actions
1. Verify the configuration of the concentration measurement application.
2. Contact customer service.
17.13.38 [121] Extrapolation Alert (Concentration)
Alert
Process Aberration
Cause
The line density or line temperature is outside the range of the concentration matrix plus the configured
extrapolation limit.
Recommended actions
1. Ensure that process density and temperature are within the limits of the concentration matrix.
2. Verify the configuration of the concentration measurement application.
17.13.39 [122] Discrete Output 3 Fixed
Alert
Output Fixed
Cause
The output is set to a constant state. A loop test may be in progress.