17.13.29 [111] Frequency Output 1 Fixed
Alert
Output Fixed
Cause
The output is configured to send a constant value. A loop test may be in progress.
Recommended actions
Check to see if a loop test is in process (the output is fixed).
17.13.30 [113] mA Output 2 Saturated
Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the settings of
Upper Range Value
and
Lower Range Value
.
Configure Lower Range Value (LRV) and Upper Range Value (URV) for the mA Output
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
17.13.31 [114] mA Output 2 Fixed
Alert
Output Fixed
Cause
The output is configured to send a constant value. A loop test may be in progress.
Recommended actions
1. Check to see if a loop test is in process (the output is fixed).
2. Exit the mA Output trim, if applicable.
17.13.32 [115] External Input Error
Alert
Process Aberration
Configuration and Use Manual
Troubleshooting
MMI-20025166
March 2019
Configuration and Use Manual
273