WARNING
If the transmitter is in a hazardous area, do not use this method to load matrices. This method requires
opening the transmitter's wiring compartment while the transmitter is powered up, and this can cause an
explosion. If the transmitter is in a hazardous area, you must use another method to load matrices.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The
transmitter's SD card and the ProLink III installation include a set of standard concentration matrices. Other
matrices are available from Micro Motion.
Each concentration matrix file must be in .matrix format.
Tip
•
If you have a custom matrix on another device, you can save it to a file, then load it to the current device.
•
If you have a matrix file in a different format, you can load it using ProLink III.
The .matrix files must be copied to the root directory of a USB drive.
You must know the derived variable that the matrix is designed to calculate.
Important
•
All concentration matrices on your transmitter must use the same derived variable.
•
If you change the setting of
Derived Variable
, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the SD card. Set
Derived Variable
before loading concentration
matrices.
Procedure
1. Choose
Menu
>
Configuration
>
Process Measurement
>
Concentration Measurement
>
Configure
Application
and ensure that the setting of
Derived Variable
matches the derived variable used by your
matrix. If it does not, change it as required and click
Apply
.
Important
If you change the setting of
Derived Variable
, all existing concentration matrices will be deleted from
the six slots, but not from the transmitter's SD card. Verify the setting of
Derived Variable
before
continuing.
2. Load the matrix.
a) Remove the cover from the transmitter's wiring compartment, open the snap flap to access the
service port, and insert the USB drive into the service port.
b) Choose
Menu
>
USB Options
>
USB Drive --> Transmitter
>
Upload Configuration File
.
c) Set
Config File Type
to Concentration Measurement Matrix.
d) Select the .matrix file that you want to load, and wait for the transfer to complete.
3. Choose Yes or No when you are asked if you want to apply the settings.
Configure process measurement applications
Configuration and Use Manual
March 2019
MMI-20025166
74
Micro Motion Model 5700 Transmitters with Configurable Outputs