17.13.20 [034] Smart Meter Verification Failed
Alert
Function Check Failed or Smart Meter Verification Aborted
Cause
The current SMV value is statistically different than the factory baseline value.
Recommended actions
Minimize process instability and repeat the test.
17.13.21 [035] Smart Meter Verification Aborted
Alert
Function Check Failed or Smart Meter Verification Aborted
Cause
The SMV test did not complete, possibly because it was manually aborted or because process conditions were
too unstable.
Recommended actions
1. Minimize process instability and repeat the test.
2. Check the SMV abort code and take appropriate steps.
3. Contact customer service.
17.13.22 [100] mA Output 1 Saturated
Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the settings of
Upper Range Value
and
Lower Range Value
.
Configure Lower Range Value (LRV) and Upper Range Value (URV) for the mA Output
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
17.13.23 [101] mA Output 1 Fixed
Alert
Output Fixed
Troubleshooting
Configuration and Use Manual
March 2019
MMI-20025166
270
Micro Motion Model 5700 Transmitters with Configurable Outputs