Process variable
Label
Display
ProLink III
FC
Concentration measurement
Density at reference
Referred Density
Density at Reference
Temperature
Dens at Ref (CM)
Specific gravity
Specific Gravity
Density (Fixed SG Units)
Spec Gravity (CM)
Standard volume flow rate
Standard Vol Flow
Volume Flow Rate at
Reference Temperature
Stnadard Volume Flow Rate
Net mass flow rate
Net Mass Flow
Net Mass Flow Rate
Net Mass Flow (CM)
Net volume flow rate
Net Volume Flow Rate
Net Volume Flow Rate
Net Volume Flow Rate (CM)
Concentration
Concentration
Concentration
Concentration (CM)
Baume
Baume
Baume
Baume (CM)
Batching
% of batch delivered
% of Fill
% of Fill Delivered
% of Fill Delivered
Energy measurement
Energy flow
Energy Flow
Energy Flow
Calorific value
Calorific Value
Calorific Value
8.2.2 Configure Lower Range Value (LRV) and Upper Range Value
(URV) for the mA Output
Display
Menu
>
Configuration
>
Inputs/Outputs
>
Channel
x
>
I/O Settings
>
Lower Range Value
Menu
>
Configuration
>
Inputs/Outputs
>
Channel
x
>
I/O Settings
>
Upper Range Value
ProLink III
Device Tools
>
Configuration
>
I/O
>
Outputs
>
mA Output
>
Lower Range Value
Device Tools
>
Configuration
>
I/O
>
Outputs
>
mA Output
>
Upper Range Value
Field Communicator
Configure
>
Manual Setup
>
Inputs/Outputs
>
Channel
x
>
mA Output
x
>
mA Output
x
Settings
>
Lower
Range Value
Configure
>
Manual Setup
>
Inputs/Outputs
>
Channel
x
>
mA Output
x
>
mA Output
x
Settings
>
Upper
Range Value
The
Lower Range Value
(LRV) and
Upper Range Value
(URV) are used to scale the mA Output, that is, to
define the relationship between
mA Output Process Variable
and the mA Output signal.
LRV
is the value of
mA Output Source
represented by an output of 4 mA.
URV
is the value of
mA Output
Source
represented by an output of 20 mA. Between
LRV
and
URV
, the mA Output is linear with the process
variable. If the process variable drops below
LRV
or rises above
URV
, the transmitter posts an output
saturation alert.
Procedure
Set
LRV
and
URV
as desired.
Enter
LRV
and
URV
in the measurement units used for
mA Output Source
.
•
Defaults: Specific to each process variable
Configuration and Use Manual
Integrate the meter with the control system
MMI-20025166
March 2019
Configuration and Use Manual
139