Recommended actions
1. Reboot or power-cycle the transmitter.
2. Retry the procedure.
3. Contact customer service.
17.13.65 Verification of mA Output 1 Failed
Alert
Electronics Failed
Cause
The reading of the mA Input does not match the reading of mA Output 1.
Recommended actions
If the alert persists, replace the transmitter.
17.13.66 Watchdog Error
Alert
Electronics Failed
Cause
The watchdog timer has expired.
Recommended actions
1. Reboot or power-cycle the transmitter to see if the alert clears.
2. If the alert persists, replace the transmitter.
3. Contact customer support.
17.14 Perform a core processor resistance test
This procedure measures the resistance between the core processor terminals in the transmitter junction box.
The procedure applies only to 4-wire remote installations and remote core processor with remote transmitter
installations.
Note
Although you can perform the same test on the terminals at the core processor, the transmitter junction box
is typically easier to access.
Procedure
1. Power down the transmitter.
2. Remove the cover of the junction box on the transmitter to access the core processor terminals.
Configuration and Use Manual
Troubleshooting
MMI-20025166
March 2019
Configuration and Use Manual
283