Instructions for use Fabius plus SW 3.n
93
Operation
O
2
flush
The O
2
flush is used for flushing and quickly filling
the breathing system and breathing bag with O
2
while bypassing the vaporizer.
During this an unmetered flow of at least 35 L/min
is given to the breathing system and the breathing
bag.
Press the
(A) key. O
2
flows for as long as
the key is held down.
Use of the O
2
flush can increase airway pressure
very quickly and abruptly change the gas
concentration.
20960
NOTE
In
ManSpont
mode, the pressure can increase
quickly and thus trigger the APL valve.
+
A