Instructions for use Fabius plus SW 3.n
67
Assembly and preparation
Inserting the flow sensor
1
Screw out the expiratory port (A).
2
Remove the flow-sensor guard
*
(B).
3
Insert the flow sensor (C).
4
Replace the flow-sensor guard* (B).
5
Screw the expiratory port (C) back on.
Connecting the exhaust port
Screw on the exhaust port (A) in the compact
breathing system from below.
WARNING
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits
that were not removed during reprocessing
can ignite when the flow sensor is in use.
– Ensure particle-free cleaning and
disinfection.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor, check for
visible damage and soiling such as
residual mucus, medication aerosols, and
particles.
– Replace flow sensors when damaged,
soiled, or not particle-free.
20967
*
optional
A
B
C
20968
A