Instructions for use Fabius plus SW 3.n
203
Technical data
Breathing system
Volume with reusable CO
2
absorber
(including absorber volume, measured in ManSpont)
Filled, without hoses
typically 4000 mL + volume of the breathing bag
Volume with Drägersorb CLIC adapter
(including absorber volume, measured in ManSpont)
Filled, without hoses
typically 3700 mL + volume of the breathing bag
Absorber volume
Reusable CO
2
absorber, filled
1500 mL
Disposable CO
2
absorber CLIC absorber Free
1200 mL
Disposable CO
2
absorber CLIC Absorber 800 Free
1200 mL
Compliance
including ventilator hose (without breathing hoses)
0.8 mL/cmH
2
O (0.8 mL/hPa)
Flexible arm for breathing bag (optional)
Volume
0.13 L
Compliance
0.13 mL/cmH
2
O (0.13 mL/hPa)
Rigid arm for breathing bag (optional)
Volume
0.11 L
Compliance
0.11 mL/cmH
2
O (0.11 mL/hPa)
Resistance
In accordance with ISO 80601-2-13, dry, with adult
breathing hose set M30146
1)
Inspiratory: Expira-
tory:
-4.7 cmH
2
O (-4.7 hPa)
4.4 cmH
2
O (4.4 hPa)
In accordance with ISO 80601-2-13, dry, without hos-
es
1)
Inspiratory: Expira-
tory:
-3.7 cmH
2
O (-3.7 hPa)
3.7 cmH
2
O (3.7 hPa)
Typical leakage
<50 mL/min