Instructions for use Fabius plus SW 3.n
87
Getting started
Set the On/Off switch (A) to the
position.
After switching on, the anesthesia workstation
starts as follows:
– A system test is performed that checks various
components. The entire results of the tests, as
well as the test results for each component, are
displayed on the screen.
– Passed:
Pass
– Failed:
Fail
– During the system test, 2 test tones are
sounded to test the speakers. To hear these
tones, do not stand any further than 4 meters
(13 feet) from the device.
– After completion of the system test, the default
settings for ventilation are loaded.
25413
A
CAUTION
Malfunction of the device function
The user must check whether the test tones were
actually sounded. The device only checks
whether the speakers are connected. A complete
loss of the ventilation function and monitoring
function would possibly not be noticed if the
speakers failed.
If not tone or only one tone is sounded, the
device is only conditionally ready for operation.
Contact DrägerService.