Instructions for use Fabius plus SW 3.n
51
Assembly and preparation
Before first operation
Activating the battery
To prevent discharging of the battery during
transport and during storage, the fuse of the
battery is not connected with the device.
1
Remove the packaging of the battery fuse.
2
Remove the fuse holder.
3
Insert fuse (A) in the fuse holder.
4
Screw the fuse holder in securely with a
quarter turn clockwise.
WARNING
Risk of device malfunction
If the battery is not sufficiently charged and
the mains power supply fails, operation
cannot be maintained long enough.
Before first operation or after storage, charge
the battery for at least 8 hours.
WARNING
Risk due to reduced power supply from the
internal battery
Batteries are wear parts. The capacity of the
battery diminishes with the period of use.
Check the functional state of the battery by
performing preventive maintenance on a
regular basis.
25413
A