Technical data
192
Instructions for use Fabius plus SW 3.n
Device data
Medical gas supply through central gas supply
Pressure range on device connection
O
2
, N
2
O, Air
41 to 87 psi (2.8 to 6 kPa x 100)
Note: Pressure fluctuations in the central gas sup-
ply must not exceed ±10 %
Gas supply connection
NIST or DISS (if required)
(Every gas inlet is equipped with a non-return valve)
Accuracy of the pressure displays
±2 % of the maximum scale value + 4 % of dis-
played value
Medical gas supply from O
2
and N
2
O cylinders
(with NIST scew connections)
Pressure at the device connection
O
2
, N
2
O
73 psi (5 kPa x 100)
(Every gas inlet is equipped with a non-return valve)
Medical gas supply from O
2
, O
2
and N
2
O or O
2
and Air cylinders
(with pin-index connections)
Cylinder connection
Pin-index hanger yokes (CGA V-1-1994)
Gas cylinder pressure
O
2
, Air
1900 psi (131 kPa x 100)
(normal full load at 21 °C, 70 °F)
N
2
O
745 psi (51.3 kPa x 100)
Pressure gauge for gas cylinders
In accordance with ASME B40.1 grade B
Pressure measurement range for gas cylinders
O
2
0 to 3000 psi (206.8 kPa x 100)
N
2
O
0 to 3000 psi (206.8 kPa x 100)
Air
0 to 3000 psi (206.8 kPa x 100)
Medical gas supply at device inlet
Dew point
>5 °C (41 °F) at ambient temperature
Oil content
<0.1 mg/m
Particles
Dust-free air (filtered with pore size < 1 µm)