Assembly and preparation
64
Instructions for use Fabius plus SW 3.n
Disposable CO
2
absorber with Drägersorb
CLIC (optional)
As an alternative to reusable CO
2
absorbers, the
disposable CO
2
absorber may also be used.
CLIC adapter allows the following single-use CO
2
absorbers to be used:
– CLIC Absorber 800+
– CLIC Absorber Free
– Infinity ID CLIC Absorber 800+
– Infinity ID CLIC Absorber Free
More information on the connection of Drägersorb
CLIC-Adapters can be found in the associated
instructions for use.
1
Press the button (G): The CLIC adapter flips
open.
2
Loosen the soda lime in the disposable CO
2
absorber, e.g., turn over the disposable CO
2
absorber a few times.
3
Remove the seal of the new disposable CO
2
absorber. Push the disposable CO
2
absorber in
the Clic adapter.
4
Push the disposable CO
2
absorber (H)
upwards into the CLIC adapter, until the CLIC
adapter engages.
CAUTION
Risk of patient recovering consciousness
If the CO
2
absorber is not correctly locked into
place, system leakage may occur.
Make sure that the absorber container is
connected tightly to the compact breathing
system during the ventilation.
WARNING
Risk of insufficient ventilation
If the CO
2
absorber is not correctly locked
into place, system leakage may occur. The
CO
2
absorber must audibly engage before
Fabius is switched on. This ensures that the
CO
2
absorber is included into the leakage test
and compliance test of the device.
After mounting and replacing, make sure the
CO
2
absorber is firmly locked into place.
20992
G
H