Cleaning, disinfection and sterilization
170
Instructions for use Fabius plus SW 3.n
Disassembly
Observe before disassembling
Switch off the device and accessory devices
and remove their mains plugs.
Sequence of disassembly
1
Unscrew the sample line and dispose of.
2
Remove the flow sensor cable.
3
Remove the O
2
sensor and the O
2
sensor
cable.
4
Remove the pressure measurement hose.
5
Remove the APL bypass hose and the
PEEP/P
MAX
hose.
6
Remove the water trap.
7
Disassemble the CO
2
absorber:
CLIC absorber (disposable) or
Reusable CO
2
absorber:
– Unscrew the CO
2
absorber from the
breathing system.
– Remove and dispose of the soda lime
dust filter (optional).
– Empty the CO
2
absorber.
– Remove the absorber insert from the
absorber container. Leave the inner and
outer sealing rings on the absorber
insert.
8
Remove the breathing bag.
9
Disassemble the breathing circuit and the
filters.
10
Unscrew the holder for the breathing bag.
11
Remove the compact breathing system:
– Remove the breathing system cover
(optional).
– Remove the inspiratory valve.
– Remove the expiratory valve.
– Unscrew the exhaust port.
– Unscrew the inspiratory port and the
expiratory port.
– Remove the flow sensor.
– Unscrew the APL valve.
12
Remove the ventilator parts.
WARNING
Risk of damage to breathing system
If the APL valve is not disassembled before
the breathing system is reprocessed, this can
lead to leakages in the breathing system.
Always remove the APL valve prior to
reprocessing.
CAUTION
Risk of injury due to breathing system heating
When the breathing system heating is switched
on, the bottom side of the compact breathing
system and the heating plate beneath it can
become hot.
Allow the breathing system to cool off before
removing.
NOTE
To prevent accidental penetration of soda lime
into the breathing system, do not transport the
breathing system with a filled reusable CO
2
absorber.