Technical data
198
Instructions for use Fabius plus SW 3.n
Ventilator
In compliance with
ISO 80601-2-13
Control ranges
Pressure limitation (
PMAX
)
15 to 70 cmH
2
O (resolution: 1 cmH
2
O)
(15 to 70 hPa (resolution: 1 hPa))
(setting must be at least 10 cmH
2
O (10 hPa)
above
PEEP
, in
SIMV/PS
mode, the
PMAX
set-
ting must be greater than
∆
PPS
+
PEEP
)
Tidal volume (
VT
)
20 to 1400 mL (resolution: 10 mL)
Tidal volume (
VT
)
20 to 1100 mL (resolution: 10 mL), in
SIMV/PS
mode
Respiratory rate (
Freq
)
4 to 60 bpm (resolution: 1 bpm)
(4 to 60 1/min (resolution: 1/min))
Ratio of inspiratory time to expiratory time (
TI:TE
)
4:1 to 1:4
Inspiratory pause (
TIP:TI
)
0 % to 50 % (resolution: 1%)
Positive end-expiratory pressure (
PEEP
)
0 to 20 cmH
2
O (resolution: 1 cmH
2
O)
(0 to 20 hPa (resolution: 1 hPa))
Inspiratory pressure (
PINSP
)
5 to 65 cmH
2
O (resolution: 1 cmH
2
O)
(5 to 65 hPa (resolution: 1 hPa)) (setting must be
at least 5 cmH
2
O (5 hPa) above
PEEP
)
Inspiratory flow (
Insp Flow
)
10 to 75 L/min (resolution: 1 L/min) in
Pressure
Control
mode
10 to 85 L/min (resolution: 1 L/min) in
Pressure
Support
and
SIMV/PS
modes
Support pressure (
∆
PPS
)
3 to 20 cmH
2
O (resolution: 1 cmH
2
O)
(3 to 20 hPa (resolution: 1 hPa)), in
Pressure
Support
mode
Support pressure (
∆
PPS
)
3 to 20 cmH
2
O, OFF (resolution: 1 cmH
2
O)
(3 to 20 hPa OFF (resolution: 1 hPa)), in
SIMV/PS
mode
Minimum respiratory rate for apnea ventilation (
Freq
Min
)
3 to 20 bpm (resolution: 1 bpm) and OFF
(3 to 20 1/min (resolution: 1/min) and OFF)
Trigger value (
Trigger
)
2 to 15 L/min (resolution: 1 L/min)
Inspiratory time (
TINSP
)
0.3 bis 4.0 sec.