
Technical data
208
Instructions for use Fabius plus SW 3.n
Essential performance characteristics
The essential performance features comprise:
– Supplying the anesthesia workstation with O
2
If the O
2
supply (central gas supply or gas
cylinder) fails, an alarm is issued.
– Supply of the patient with adequately
oxygenated breathing gas
If the breathing gas contains insufficient levels
of O
2
, an alarm is issued.
– Monitoring of the airway pressure and the
expiratory minute volume
Alarms are issued depending on the set alarm
limits.
– Measurement accuracy of the O
2
measurement.
Alarms are issued depending on the set alarm
limits. If the O
2
sensor fails, an alarm is issued.
Pin assignment
Pin 1
n/c
Pin 2
TXD
Pin 3
RXD
Pin 4
n/c
Pin 5
GND
Pin 6
n/c
Pin 7
n/c
Pin 8
n/c
Pin 9
n/c
NOTE
In accordance with general safety standards,
additional components are required for a
complete anesthesia workstation.