Instructions for use Fabius plus SW 3.n
101
Operation
Ventilation mode
Pressure Control
(optional)
Changing to
Pressure Control
mode
In the following examples and illustrations, the
change from
Volume Control
mode to
Pressure
Control
mode is described:
1
Press the
Pressure Control
(A) key.
2
Adjust ventilation settings (B).
3
Confirm the new mode.
Due to the influence of compliance and resistance,
the set value for
Freq Min
in
Pressure Control
mode might not exactly be applied.
In the following table, all parameters (B) in
Pressure Control
mode, together with the
corresponding setting ranges and factory settings,
are listed.
2547
2
254
73
A
B
Volume Control
Pressure Control
Parameter
Setting range Factory set-
ting
Inspiratory
pressure
PINSP
[cmH
2
O]
([hPa])
5 to 65,
min.
PEEP
+5
15
Respiratory
rate
Freq
[bpm]
([1/min])
4 to 60
12
Inspiratory
time:expiratory
time
TI:TE
4:1 to 1:4
1:2
Inspiratory flow
Insp Flow
[L/min]
10 to 75
30
PEEP
[cmH
2
O]
([hPa])
0 to 20
0