Instructions for use Fabius plus SW 3.n
151
Configuration
If the connected anesthetic gas monitor detects
desflurane, Fabius reacts as follows:
–
Des auto
is displayed in the status bar (A).
– The softkey
DesComp
ON
/
OFF
(B) is no
longer displayed.
If the communication to the anesthetic gas monitor
fails, Fabius reacts as follows:
– The automatic desflurane compensation is
switched off.
– The message
Des auto
in the status bar (A) is
no longer displayed.
– The softkey
DesComp
OFF
(B) is displayed.
To switch the desflurane compensation back on:
Press the
DesComp
OFF
softkey (B).
The label of the softkey changes to
DesComp
ON
.
Accessing the alarm logbook
1
Press the
AccessAlarm
Log
softkey (A).
2
To scroll through the alarm logbook (D), turn
the rotary knob.
Clearing the alarm logbook
Select
Clear Alarm Log
(C) and confirm.
Closing the alarm logbook
Select the input arrow (B) and confirm.
The pressure waveform and softkeys are
displayed again.