Assembly and preparation
78
Instructions for use Fabius plus SW 3.n
Inserting a new O
2
sensor capsule
1
Remove the O
2
sensor housing from the
inspiratory valve.
2
Unscrew the screw cap (A) from the O
2
sensor
housing.
3
Take the new O
2
sensor capsule from the
package.
4
Place the O
2
sensor capsule (B) in the housing
so that the ring-shaped conductor touches the
contacts in the housing.
5
Screw on the screw cap (A) tightly by hand.
6
Reinsert the O
2
sensor housing in the
inspiratory valve.
WARNING
Risk of electric shock
If the O
2
sensor is replaced during operation,
it can lead to transferring of leakage current.
Do not touch the patient.
20971
WARNING
Danger of erroneous O
2
measurement.
An incorrectly mounted O
2
sensor will lead to
incorrect measurement results.
Make sure that the O
2
sensor is inserted
correctly in the inspiratory valve, see page 22.
A
B