Troubleshooting
162
Instructions for use Fabius plus SW 3.n
(23/31)
APNOEA FLOW !!!
The apnea flow alarm based
on a time staggering.
In the
Volume Control
,
Pressure Control
,
SIMV/PS
modes with
Freq
≥
6 or in
Pressure Support
mode
with apnea ventilation deacti-
vated:
Warning =
VT
<20 mL for >30
seconds
In the
ManSpont
,
SIMV/PS
modes with
Freq
< 6 or in
Pressure Support
mode
with apnea ventilation acti-
vated:
Warning =
VT
<20 mL for >60
seconds
Breathing/ventilation
stopped.
Check ventilator.
Leakage or disconnection in
the breathing system.
Check breathing system.
Alarm
priority
Alarm
Cause
Remedy