Instructions for use Fabius plus SW 3.n
133
Configuration
The test results (B) are displayed on the screen.
After completion of the system test, the total test
results (C) are displayed, see chapter ''Checking
the readiness for operation'' on page 88. If the
system test was successful, the
Standby
mode is
activated.
Calibrating the flow sensor
1
Press the
Calibrate
Flow
Sensor
softkey (A).
2
Follow the instructions on the screen.
At the start of the calibration, the instructions are
hidden and the following message is displayed
above the standby softkeys (B):
Flow Calibration in progress
After the calibration, one of the following two
messages are displayed above the standby
softkeys (B):
Flow Calibration completed - reconnect
expiratory hose
or
Flow Calibration Failed
Troubleshooting with failed flow calibration
Repeat the calibration.
Replace the flow sensor.
If the calibration continues to fail, contact
DrägerService or the authorized local service
partner.
25494
25463
B
C
A
B
Standby