Instructions for use Fabius plus SW 3.n
149
Configuration
Calibrating the O
2
sensor
1
Press the
Calibrate
O
2
Sensor
softkey (A).
2
Follow the instructions on the screen.
To calibrate the O
2
sensor, proceed as described
in chapter ''Calibrating the O
2
sensor'' on page
134.
During the calibration, the O
2
value in the window
(B) of the O
2
monitoring is replaced by the word
CAL
. The calibration time is approx. 15 seconds.
After successful calibration, the O
2
measured
value is again displayed.
If the calibration was not successful, replace
the O
2
sensor capsule in the O
2
sensor
housing, see chapter ''Inserting a new O
2
sensor capsule'' on page 78.
If the calibration continues to fail, contact
DrägerService or the authorized local service
partner.
25505
25506
A
Volume Control
B
Volume Control