Annex
224
Instructions for use Fabius plus SW 3.n
If any one of the test points is not passed, the
device must not be used. Contact DrägerService
or the authorized local service partner.
P
Before connecting to the patient
– All vaporizers are switched off (the control
dials are in the
0
position).
– The APL valve is set to the desired pres-
sure.
– All fresh gas flow indicators show
0
.
– The scavenging flow of the endotracheal
suction is present.
– The breathing system is ready for opera-
tion (the breathing bag is correctly posi-
tioned and all hoses are correctly
connected).
– CO
2
absorber is present on the device and
is adequately filled.