Operation
92
Instructions for use Fabius plus SW 3.n
Setting the anesthetic gas concentration
Prerequisite: The vaporizer is mounted in
accordance with the instructions for use of the
respective vaporizer.
The following section describes the operation of
Vapor 2000.
Set the fresh-gas flow on the anesthesia
workstation.
If the control dial is in position T:
1
Press the 0 key (A) and engage the control
dial (B) to position 0. To ensure pressure
equalization, wait at least 15 seconds.
2
Press the 0 key (A) and set the control dial (B)
counterclockwise to the desired anesthetic gas
concentration.
3
Check the filling level on the sight glass (C)
regularly. The filling level must be between
minimum and maximum.
4
When the refill mark (D) is reached, 250 mL
(normal anesthetic agent bottle) can be refilled.
5
Fill the vaporizer at the latest when the
minimum mark (E) is reached, refer to
instructions for use of the respective vaporizer.
20964
WARNING
Risk of patient injury
When the vaporizer is in the control dial
position T, has heated up due to high ambient
temperature, and is then used, a high
anesthetic gas concentration might be
delivered.
After connecting the vaporizer to the
anesthesia workstation, always turn the
control dial of the vaporizer to position 0 and
wait at least 15 seconds to enable a pressure
equalization.