Instructions for use Fabius plus SW 3.n
127
Monitoring
Volume alarms
Volume alarms in automatic ventilation modes
If the volume alarm messages are activated and
Fabius does not detect a breath in a specific time
period, the alarm
APNOEA FLOW !!
or
APNOEA
FLOW !!!
is triggered, see chapter ''Alarm – Cause
– Remedy'' on page 160.
Volume alarms in
ManSpont
If the volume alarm messages are activated and
Fabius does not detect a breath in a specific time
period, after 30 seconds the alarm
APNOEA
FLOW !!
with the priority CAUTION is triggered. If
this alarm is not remedied, the priority increases
after a further 30 seconds to WARNING.
The volume alarm messages are automatically
activated when changing from
Standby
mode in a
ventilation mode.
Setting the minute volume alarm limits
The standard alarm limits configured for the
ventilation mode can be used unchanged, see
chapter ''Changing the alarm limits'' on page 141,
or
the alarm limits can be set individually for the
current case.
1
Press the
(A) key.
The dialog window (B) with the alarm limits opens.
2
Adjust upper and lower alarm limit values of the
minute volume (
MV
), see setting ranges of the
alarm parameters in chapter ''Changing the
alarm limits'' on page 141.
3
Confirm new values.
Deactivating the volume alarms
The volume alarms can be switched on and off
during operation by pressing the
key (A), see
chapter ''Switching the volume alarms on and off''
on page 148.